Gw-instek PEL-2000 Series Programming User Manual Instrukcja Użytkownika Strona 91

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 170
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 90
COMMAND DETAILS
91
:OCP:CHANnel:DELay
Channel Specific
Command
Description
Queries or sets the test delay time for the OCP Test
Automation function.
Syntax
:OCP:CHANnel:DELay {<NRf>[ S ] | MIN | MAX}
Parameter
<NRF>[S]
The delay time in seconds (5mS~160S).
MAX
The maximum delay time.
MIN
The minimum delay time.
Example
:OCP:CHAN:DEL MAX
Set the delay time to the
maximum value.
Query Syntax
:OCP:CHANnel:DELay?
Return Parameter
<NR2>
Returns the delay time in seconds.
Query Example
:OCP:CHAN:DEL?
5.0
Returns the delay time.
:OCP:CHANnel:TRIGger
Channel Specific
Command
Description
Queries or sets the voltage trigger for when the
power supply OCP has been triggered.
Syntax
:OCP:CHANnel:TRIGger {<NRf>[ V ] | MIN | MAX}
Parameter
<NRF>[V]
The trigger voltage level.
MAX
The maximum trigger voltage.
MIN
The minimum trigger voltage.
Example
:OCP:CHAN:TRIG MAX
Set the trigger voltage
level to the maximum
value.
Query Syntax
:OCP:CHANnel:TRIGger?
Return Parameter
<NR2>
Returns the trigger voltage level in volts.
Query Example
:OCP:CHAN:TRIG?
5.0
Returns the trigger level.