COMMAND DETAILS
149
SEQuence Subsystem
:SEQuence:EDIT:POINt ............................................... 149
:SEQuence:END ........................................................... 150
:SEQuence:POINt:RESistance .................................... 150
:SEQuence:POINt:CURRent ........................................ 151
:SEQuence:POINt:RISE/FALL ..................................... 151
:SEQuence:POINt:TIME .............................................. 152
:SEQuence:REPeat ....................................................... 153
:SEQuence:END:LOAD................................................ 153
:SEQuence:VOLTage:RANGe ...................................... 154
:SEQuence:LOOP:STARt ............................................. 155
:SEQuence:CHANnel:TIME ......................................... 155
:SEQuence:RUN ........................................................... 156
:SEQuence:SAVE .......................................................... 156
:SEQuence:TRIGger:IN ................................................ 156
:SEQuence:TRIGger:OUT ............................................ 157
:SEQuence:TRIGger:IN:CHANnel .............................. 157
:SEQuence:EDIT:POINt
Channel Specific
Command
Description
Sets the current point in the sequence. The
SEQuence:END command should first be used to
set the number of points.
Syntax
:SEQuence:EDIT:POINt <NR1>
Parameter
<NR1>
Points
1~last point
1~ last point.
Example
:SEQ:EDIT:POIN 3
Sets the point to number
3.
Query Syntax
:SEQuence:EDIT:POINt?
Return Parameter
<NR1>
Points
1~last point
Returns the current point.