STATUS REGISTERS
165
The Go/NoGo bit is set when Go/NoGo
limits have been exceeded, when
Go/NoGo SPEC has been enabled.
If the under voltage condition occurs the
UVP bit is set.
The condition register indicates the status of the
electronic load. The condition register can only be
changed by a change in the condition of the
electronic load. Reading the condition register does
not change the state of the condition register.
The PTR/NTR (Positive/Negative transition)
register determines the type of transition
conditions that will trigger an event. Only the
Channel Status Register and Questionable Status
Register can be transition programmed
The Event Register indicates if an event has been
triggered according to the transition settings from
the PTR/NTR Register.
The Enable register determines which status
event(s) are enabled. Any status events (OC, OV,
OP, RV, OT, G/N, UVP) that are enabled will set
the corresponding channel bit in the Channel
Summary Event Register.
Channel Summary
The Channel Summary Registers consolidate the
channel status of all 4/8 channels, depending on
the electronic load.