PEL-2000 Programming Manual
44
:CONFigure:PROTection:CURRent:LEVel
Channel Specific
Command
Description
Sets the current protection level for the
current/specific channel. The level can be set to
any applicable level or to the channel
maximum/minimum.
Syntax
:CONFigure:PROTection:CURRent:LEVel
<NRf>[A]|MIN|MAX
Parameter
<NRf>
Current Protection Level
.3
300mA
0.3A
300mA
300MA
300mA
MIN
Sets to the minimum level
MAX
Sets the current limit to
the maximum level
Example
:CONF:PROT:CURR:LEV
MAX
Sets the current limit to
20.40A (PEL2020)
Query Syntax
: CONFigure:PROTection:CURRent:LEVel? [MIN|MAX]
Return Parameter
<NRf> 1 unit = 1 amp
Current protection level
1
1 amp.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:CONF:PROT:CURR:LEV?
0.30
Current protection level
is at 300mA.