Gw-instek PEL-2000 Series Programming User Manual Instrukcja Użytkownika Strona 39

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 170
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 38
COMMAND DETAILS
39
Example
:CONF:VOLT:ON 30MV
Set Von to 30 millivolts.
Query Syntax
:CONFigure:VOLTage:ON?
Return Parameter
<NR2> I unit = I volt
Von value (volts)
1
1 volts
Query Example
:CONF:VOLT:ON?
0.03
Von is set as 30 millivolts
(0.03 volts).
:CONFigure:VOLTage:RANGe
Channel Specific
Command
Description
Sets Voltage range for CC mode.
Syntax
:CONFigure:VOLTage:RANGe <NRf>[V]|L|H
Parameter
<NRf>[ V] , L, H
Range
16
Low range*
80V
High range*
L
Low range
H
High range
*Load module dependent, PEL-2020 shown.
Example
:CONF:VOLT:RANG L
Sets the range to Low for
the channel.
Query Syntax
:CONFigure:VOLTage:RANGe?
Return Parameter
<NR2>
Range
16
Low PEL-2020,2030,2040
125
Low PEL-2041
80
High PEL-2020,2030,2040
500
High PEL-2041
Query Example
:CONF:VOLT:RANG?
500
Returns the voltage
range. In this case high
for the PEL-2041.
Przeglądanie stron 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 169 170

Komentarze do niniejszej Instrukcji

Brak uwag