Gw-instek PEL-2000 Series Programming User Manual Instrukcja Użytkownika Strona 127

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 170
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 126
COMMAND DETAILS
127
Query Example
:STAT:QUES:COND?
2
OV (overvoltage) error.
:STATus:QUEStionable:ENABle
Channel Specific
Command
Description
Sets which events are enabled in the Questionable
Status Enable register. The mask values are the bit
weights of the events. See page 166 for details.
Syntax
:STATus:QUEStionable:ENABle <NR1>
Parameter
<NR1>
Event
<NR1>
Event
1
OC
16
OT
2
OV
32
G/N
4
OP
64
UVP
8
RV
128~65535
Not used
Example
:STAT:QUES:ENAB 12
Events OP (Bit 3) and RV
(Bit 4) are enabled in the
Questionable Status
Enable register.
Query Syntax
:STATus:QUEStionable:ENABle? <NR1>
Return Parameter
<NR1>
Event
<NR1>
Event
1
OC
16
OT
2
OV
32
G/N
4
OP
64
UVP
8
RV
128~65535
Not used
Query Example
:STAT:QUES:ENAB?
4
The OP event is enabled.
:STATus:QUEStionable[:EVENt]?
Channel Specific
Command
Description
Returns the status of the Questionable Status Event
register. The Questionable Status Event register is
cleared upon reading.
Przeglądanie stron 126
1 2 ... 122 123 124 125 126 127 128 129 130 131 132 ... 169 170

Komentarze do niniejszej Instrukcji

Brak uwag