Gw-instek PEL-2000 Series Programming User Manual Instrukcja Użytkownika Strona 60

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 170
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 59
PEL-2000 Programming Manual
60
:CURRent:STATic:LOW:RISE/FALL
Channel Specific
Command
Description
Sets the low range rising/falling slew rates.
Syntax
:CURRent:STATic:LOW:RISE/FALL <NRf+>[A/uS]
Parameter
<NRf+>[A/uS]
Slew Rate
RISE/FALL 0.078A/uS
Sets the rising/falling slew
rate to 0.078A/uS
RISE/FALL 1
Sets the rising/falling slew
rate to 1A/uS
RISE/FALL MIN
Sets to the slowest
rising/falling slew rate.
RISE/FALL MAX
Sets to the fastest
rising/falling slew rate.
Example
:CURR:STAT:LOW:RISE .001
Sets the rising slew rate
to 0.001A/uS.
Query Syntax
: CURRent:STATic:LOW:RISE/FALL? [MIN|MAX]
Return Parameter
<NR2> [MAX|MIN]
Slew rate
1 Unit=1 amp/uS
Returns the slew rate.
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:CURR:STAT:LOW:RISE?
MIN
0.078
For low range CC mode,
the Minimum value for
the rising slew rate is
0.078 A/uS for the
specific channel.
:CURRent:STATic:HIGH:AVALue/BVALue
Channel Specific
Command
Description
Sets the high range A/B Value for constant current
static mode.
Syntax
:CURRent:STATic:HIGH:AVALue/BVALue <NRf+>[A]
Parameter
NRf+[A]
Przeglądanie stron 59
1 2 ... 55 56 57 58 59 60 61 62 63 64 65 ... 169 170

Komentarze do niniejszej Instrukcji

Brak uwag