68
Syntax
:CURRent:DYNamic:HIGH:L1/L2 <NRf+>[A]
Parameter
NRf+[A]
L1 10
Sets L1 to 10 Amps. (High
range only)
L2 20
Sets L2 to 20 Amps. (High
Range only)
L1/L2 MIN
Sets L1 or L2 to the
minimum level for the
specific channel.
L1/L2 MAX
Sets L1 or L2 to the
maximum Level for the
specific channel.
Example
:CURR:DYN:HIGH:L1 10
In high range CC
dynamic mode, Set L1
(level 1) to 10 amps.
Query Syntax
:CURRent:DYNamic:HIGH:L1/L2? [MIN|MAX]
Return Parameter
<NR2> [MAX|MIN]
Return value
1 unit= 1 amp
Returns the current of
Level 1/ 2 (L1/L2).
MAX/MIN
Returns the allowable
maximum and minimum.
Query Example
:CURR:DYN:HIGH:L2?
MAX
20.4000
Returns the maximum
current allowed for the
channel. (PEL-2020)
:CURRent:DYNamic:HIGH:RISE/FALL
Channel Specific
Command
Description
Sets the high range rising/falling slew rate for CC
dynamic mode for the specific channel.
Syntax
:CURRent:DYNamic:HIGH:RISE/FALL <NRf+>[A/uS]
Parameter
<NRf+>[A/uS]
Slew Rate
RISE/FALL 0.8A/uS
Sets the rising/falling slew
rate to 0.8A/uS