PEL-2000 Programming Manual
76
Parameter
ON/1
Shorting is On
OFF/0
Shorting is Off
Example
:LOAD:SHOR ON
Short circuits the load
module channel.
Query Syntax
:LOAD:SHORt[:STATe]?
Return Parameter
<NR1>
Short Load module
1
Shorting is activated
0
Shorting is deactivated
Query Example
:LOAD:SHOR?
0
Shorting is deactivated
on the specific channel.
:LOAD:SHORt:KEY
Channel Specific
Command
Description
The SHORT key can be set to Toggle or Hold
mode.
Syntax
:LOAD:SHORt:KEY {TOGGLE|1|HOLD|0}
Parameter
TOGGLE/1
Sets the SHORT key to
toggle mode
HOLD/0
Sets the SHORT key to
hold mode
Example
:LOAD:SHOR:KEY 1
Set the SHORT key to
toggle.
Query Syntax
:LOAD:SHORt:KEY?
Return Parameter
<NR1>
Mode
1
Toggle mode is active
0
Hold mode is active
Query Example
:LOAD:SHOR:KEY?
0
Hold mode is active for
the specific channel.