PEL-2000 Programming Manual
120
:SPECification:VOLTage:H/L/C
Channel Specific
Command
Description
Sets the high(H), low(L) and center(C) Go/NoGo
voltage limit specifications. Applicable to CC and
CR mode only.
Syntax
:SPECification:VOLTage:H/L/C <NRf+>[V]
Parameter
<NRf+>[V] 1 unit = 1 volt
Go/NoGo voltage limit
1
1 volt
1V
1 volt
Example
:SPEC:VOLT:H 2V
Sets the Go/NoGo high
voltage limit to 2 volts.
Query Syntax
:SPECification:VOLTage:H/L/C? <NR2>
Return Parameter
<NR2>
Go/NoGo voltage
1 unit = 1 volt
Returns the limit voltage
Query Example
:SPEC:VOLT:H?
2.000
The voltage limit is 2
volts.
:SPECification:CURRent:H/L/C
Channel Specific
Command
Description
Sets the high(H), low(L) and center(C) Go/NoGo
current limit specifications. Applicable to CV
mode only.
Syntax
:SPECification:CURRent::H/L/C <NRf+>[A]
Parameter
<NRf+>[A] 1 unit = 1 amp
Go/NoGo current limit
1
1 amp
1A
1 amp
Example
:SPEC:CURR:H 1A
Sets the Go/NoGo high
current limit to 1 amp.
Query Syntax
:SPECification:CURRent:H/L/C? <NR2>