PEL-3000 Programming Manual
94
Related
Commands
:POWer:LEVel
Parameter
<NRf>[ W ]
“Set” power value
MINIMUM or MIN
Minimum power level
MAXIMUM or MAX
Maximum power level
Return parameter
<NR2>
Returns the “Set” power
value.
Example
:POWer:SET MIN
Sets the power value to the minimum.
Query example
:POWer:SET?
>10W
Returns the “Set” power setting.
:POWer:LEVel
Set
Query
Description
Sets or queries the CP mode % level (percentage of
the Set power value) when the dynamic mode
units are set to Percent.
Syntax
:POWer:LEVel {<NRf> | MINimum | MAXimum }
Query Syntax
:POWer:LEVel?
Related
Commands
:POWer:Set
Parameter
<NRf>
“% level” power value
MINIMUM or MIN
Minimum % power level
MAXIMUM or MAX
Maximum % power level
Return parameter
<NR2>
Returns the “% level”
current value.
Example
:POWer:LEVel MIN
Sets the % level power value to the minimum.
Query example
:POWer:LEVel?
>50
Returns the “% Level” as 50 percent.