COMMNAND OVERVIEW
141
:STATus:PRESet
Set
Description
The :STATus:PRESet command resets the enable
registers NTR/PTR registers from Operation
Status, Questionable Status and CSummary Status
registers.
Preset Value
Register
Preset
Operation Status Enable
All bits set to 1
Operation Status PTR
All bits set to 1
Operation Status NTR
All bits set to 0
Questionable Status Enable
All bits set to 0
Questionable Status PTR
All bits set to 1
Questionable Status NTR
All bits set to 0
CSummary Status Enable
All bits set to 0
CSummary Status PTR
All bits set to 1
CSummary Status NTR
All bits set to 0
Syntax
:STATus:PRESet
:STATus:QUEStionable:CONDition?
Set
Query
Description
Returns the status of the Questionable Status
Condition register. See page 154 for details.
Query Syntax
:STATus:QUEStionable:CONDition?
Return parameter
<NR1>
Condition
<NR1>
Condition
1
OV
512
UV
2
OC
1024
EXT
8
OP
2048
REV
16
OT
All other
bits
Not used
Query example
:STATus:QUEStionable:CONDition?
>1
Indicates an over voltage (OV) error has occurred.