COMMNAND OVERVIEW
59
Return parameter
<ASCII string>
Returns the range and the
step resolution.
Example
:STEP:CRH 0.8
Sets the step resolution to 0.8.
Query example
:STEP:CRH?
>CRH:0.8
Returns the step resolution (0.8).
[:CONFigure]:STEP:CRM
Set
Query
Description
Configures the step resolution for CR Medium
Range. Note: The step resolution setting will be
automatically rounded to the closest multiple of
the base resolution.
Syntax
[:CONFigure]:STEP:CRM {<NRf>[ MHO ] | MINimum
| MAXimum }
Query Syntax
[:CONFigure]:STEP:CRM?
Parameter
<NRf>[ MHO ]
Step resolution. [MHO]
MINIMUM or MIN
Minimum step resolution
MAXIMUM or MAX
Maximum step resolution
Return parameter
<ASCII string>
Returns the range and the
step resolution.
Example
:STEP:CRM 0.08
Sets the step resolution to 0.08.
Query example
:STEP:CRM?
>CRM:0.08
Returns the step resolution (0.08).