REMOTE CONTROL
83
:CALCulate:MARKer:PEAK:DATA?
Query
Description
Returns all the top 10 peak data values in CSV
format (returns the contents of the peak table).
The <csv data> data contains 10 pairs of data from
the top 10 peaks. Each pair includes the peak
frequency and the peak amplitude. There are a
total of 10 pairs of data points (20 data entries in
total) for the <csv data> data.
Query syntax
:CALCulate:MARKer:PEAK:DATA?
Return parameter
<csv data>
pk#1 freq, pk#1 amp,….. pk#10 freq,
pk#10 amp.
Example
:CALCulate:MARKer:PEAK:DATA?
>1.250e+08,-5.052e+01,1.065000000e+09,
:CALCulate:MARKer:PEAK:EXCursion
Set
Query
Description
Sets or queries the peak excursion value.
Syntax
:CALCulate:MARKer:PEAK:EXCursion <rel ampl>
Query Syntax
:CALCulate:MARKer:PEAK:EXCursion?
Parameter
<rel ampl>
Peak excursion dB (offset from
threshold)
Return parameter
<NR3>
Peak excursion in dB.
Example
:CALC:MARK:PEAK:EXC 6 db
:CALCulate:MARKer:PEAK:SORT:TYPE
Set
Query
Description
Sets or queries the peak sort type for the peak
table.
Syntax
:CALCulate:MARKer:PEAK:SORT:TYPE
{FREQuency|AMPLitude}
Query Syntax
:CALCulate:MARKer:PEAK:SORT:TYPE?