Gw-instek GSP-930 Programming Manual Instrukcja Użytkownika Strona 165

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 195
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 164
REMOTE CONTROL
163
:SOURce:POWer[:LEVel][:IMMediate]
[:AMPLitude]:STEP
Set
Query
Description
Sets or queries the tracking generator step level.
Syntax
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]:
STEP <rel_ampl>
Query Syntax
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]:
STEP?
Parameter
<rel_ampl>
<NRf>
Return parameter
<NR3>
dB
Example
:SOUR:POW:STEP .5 db
:SOURce:POWer[:LEVel][:IMMediate]
[:AMPLitude]:STEP:AUTO
Set
Query
Description
Turns the tracking generator step level setting to
auto (on) or manual (off).
Syntax
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]:
STEP:AUTO {OFF|ON|0|1}
Query Syntax
:SOURce:POWer[:LEVel][:IMMediate][:AMPLitude]:
STEP:AUTO?
Parameter
0
Turn TG step level to manual (off).
1
Turn TG step level to auto (on).
OFF
Turn TG step level to manual (off).
ON
Turn TG step level to auto (on).
Return parameter
0
TG step level is set to manual.
1
TG step level is set to automatic.
Example
:SOUR:POW:STEP:AUTO 1
:SOURce:POWer:MODE
Set
Query
Description
Sets the Power Sweep mode.
Przeglądanie stron 164
1 2 ... 160 161 162 163 164 165 166 167 168 169 170 ... 194 195

Komentarze do niniejszej Instrukcji