OUT Volt Current Power OUT Volt Current Power
0 1 5 1 1 0 410 1 0 2 410
99 1 5 1 1 0 410 1 0 2 410
After you turn on the address to select the interface, Sequence Data will be
transferred to PSB by clicking on the button UPLOAD or DOWNLOAD.
And, we offer to storage to File by Import or Export function.
Sequence data transfer program setting sheet
Seq
No
MAIN
OUTPU
PSB Series Sequence Data Transfer Program
Duration
CHA
CHB
Figures for sequence setting list on the Sheet1
For 1 Channel model, the Output ON/OFF can be config at MAIN OUTPUT column.
The duration of step time can be config at Duration column by unit as seconds.
(Range is 1 to 65535, The Maximum time will be 18 hours.)
To make the settings for each ch in 2ch model, please set chA in 1ch model.
OUT can set as 1/0 for ON/OFF of the output.
Voltage setting Volt column, current setting Current column, power setting the Power
colume, I enter V, A, W in each.
Please check the check button of model selection and calculation range will vary
depending on the model
For User Interface
COM: (USB-COM converter is also available.) Address setting is port number of the
COM port on the PC side. Please use the inter-link cable to use the hardware flow
cable.
USB: Please use the USB-CDC Driver on the website. Address setting is port
number of the COM port on the PC side.
GPIB: Please use it to install the driver according to the instructions of the board.
Setting the address is the address of the PSB PC side. In addition, please contact us
for GPIB board of National Instruments other than.
It is not possible to sequence transferred to the PSB power of local bus connection.