130
parameter
TIME3M
TIME6M
TIME15M
TIME30M
TIME1H
TIME3H
TIME6H
Example
:DMM:TIMe 6M
Sets the measurement interval time to 6M.
:DMM:CURSORPOS
Description
Sets the cursor position for trend plot in the
Advanced DMM function.
This command is only applicable when the Hold
function is turned on.
Syntax
:DMM:CURSORPOS { <NR1> | ? }
Parameter
<NR1>
0 ~ 400
Return parameter
<string>
Returns the cursor position (#data);
time at position; measurement value
at position
Example
DMM:HOLD ON
DMM:CURSORPOS 200
:DMM:CURSORPOS?
200;20.0s;0.0000V
Returns cursor position #200 @ 20s with a value of
0V.