GDS-200 & GDS-300 Programming Manual
124
Replay Commands
The replay function automatically logs acquisitions when the scope
is in Run mode.
:REPLAY:TOTalnum .................................................. 124
:REPLAY:CURRent ..................................................... 124
:REPLAY:TOTalnum
Description
Queries the total number of waveforms that can be
stored for the replay function.
Syntax
:REPLAY:TOTalnum?
Related
commands
REPLAY:CURRent
Return parameter
<NR1>
Returns the total number of
waveforms.
Example
:REPLAY:TOTalnum?
>3000
A maximum of 3000 waveforms can be stored for
the replay function.
:REPLAY:CURRent
Description
Sets or queries the current waveform that is
displayed when using the replay function. Note:
the scope may need to be in Stop mode to use this
function.
Syntax
REPLAY:CURRent { FORward | BACKward | <NRf> | ? }
Related
commands
REPLAY:TOTalnum?
Parameter/Return
parameter
FORward
Moves to the next waveform in
memory
Query
Set
Query