Compact Digital Storage Oscilloscope and Digital Multimeter GDS-200 and GDS-300 Series USER MANUAL GW INSTEK PART NO. 82DS-22000EA1 ISO-9001 CE
GDS-200 & GDS-300 Series User Manual 10 GETTING STARTED This chapter will allow you to quickly set up a GDS-200 or GDS-300 series oscilloscope
GDS-200 & GDS-300 Series User Manual 100 Touch Sounds Background When turned on, the Touch Sounds function will produce a buzzer sound when
OPERATION 101 Steps 1. From the Utility menu, press Hardcopy and select the save function for the Hardcopy function. 2. The Hardcopy Save key ca
GDS-200 & GDS-300 Series User Manual 102 Configure USB Device Port Background The setting, “USB device port” in the Utility menu determines w
OPERATION 103 Configure Autoset Mode Background There are two operating modes for the Autoset function: Fit Screen Mode and AC Priority Mode. Fit S
GDS-200 & GDS-300 Series User Manual 104 Set Ruler Background The Ruler function adds a scale to the graticule. Note this mode only functions
OPERATION 105 Calibration Background The Calibration menu allows you to perform vertical calibration on the oscilloscope channels. Steps 1. From t
GDS-200 & GDS-300 Series User Manual 106 5. Repeat the above steps for CH2. 6. Reset the scope when the calibration has finished. Vertical
OPERATION 107 Acquire Menu The Acquire Menu sets the acquisition mode, channel mapping (XY mode), record length and the sample interpolation type.
GDS-200 & GDS-300 Series User Manual 108 Select Acquisition Mode Background The acquisition process samples the analog input signals and conv
OPERATION 109 Show Waveform in XY Mode Background The XY mode maps the input of channel 1 to the input of channel 2. This mode is useful for observ
GETTING STARTED 11 Accessories ... 14 A
GDS-200 & GDS-300 Series User Manual 110 Set the Record Length Background The number of samples that can be stored is set by the record lengt
OPERATION 111 Set the Interpolation Background The GDS-200/GDS-300 has two types of interpolation modes: ET (Equivalent Time) and Sin(x)/x interpo
GDS-200 & GDS-300 Series User Manual 112 Display Menu The Display Menu defines how the waveforms and parameters appear on the display. Access
OPERATION 113 Vectors: Both the sampled dots and the connecting line are displayed. Steps From the Display menu press Type and select the display
GDS-200 & GDS-300 Series User Manual 114 Set the Waveform Intensity Background The intensity level of a signal can also be set to mimic the i
OPERATION 115 Set the Graticule Intensity Background The graticule intensity level can also be configured. Steps From the Display menu press Grati
GDS-200 & GDS-300 Series User Manual 116 Set the Graticule Type Background There are 4 types of selectable graticule. Steps From the Display
OPERATION 117 Set the Brightness Level Background The Brightness menu sets the display brightness and the display power management, known as ECO mo
GDS-200 & GDS-300 Series User Manual 118 Go-NoGo Menu A Go-NoGo test checks if a waveform fits inside a user-specified maximum and minimum bou
OPERATION 119 Enter Sets the NoGo condition to when the input signal stays within the limit boundary. Exit Sets the NoGo condition
GDS-200 & GDS-300 Series User Manual 12 GDS-200 and GDS-300 Series Overview Series Lineup There are 6 models in the series, divided by DSO
GDS-200 & GDS-300 Series User Manual 120 Max/Min Individually sets the offset for the upper and lower templates. These are based on the grou
OPERATION 121 Save Template Saves both the upper and lower template to R1 and R2 respectively. Max/Min Example Max templateMin templateOffset from
GDS-200 & GDS-300 Series User Manual 122 Running a Go-NoGo Test Background The Go-NoGo controls are located at the bottom left-hand corner of
OPERATION 123 Results 2. When Go-NoGo is running, the Fail/Total ratio is displayed in the title bar. The first digit represents the number of fail
GDS-200 & GDS-300 Series User Manual 124 Save Operations The Save function configures the save file destination, file format and in the case
OPERATION 125 Saving and Configuring Image Files Background Image files can be saved to the internal memory in a number of file formats. Custom no
GDS-200 & GDS-300 Series User Manual 126 File Manager Goes to the file manager to set the save-file path or to create new folders. See page
OPERATION 127 Brief Mode Brief mode adds a series of user-selectable data to the bottom of an image file. It also reverts the normally black backgro
GDS-200 & GDS-300 Series User Manual 128 9. To preview the image file before saving, press the Preview button. Example Image File with Brief
OPERATION 129 Waveform tabSave button From 3. Use the From option to select the source waveform: 1 Channel 1 2 Channel 2 M Math source R1
GETTING STARTED 13 Features Simultaneous DMM and DSO operation Portrait and landscape modes Large 800 x 480 TFT panel and capacitive touch
GDS-200 & GDS-300 Series User Manual 130 5. If All Displayed waveforms are to be saved, then the following destination options are available
OPERATION 131 For example if the waveform is saved as DS001.lsf then the note will be saved as DS001_LSF.txt. Saving and Configuring Setups Backgro
GDS-200 & GDS-300 Series User Manual 132 File Pressing File will allow you to either set the file path or rename the file from the default o
OPERATION 133 Saving and Configuring Images, Waveforms and Setups All at Once Background Images, Waveforms and Setup files can all be saved at the
GDS-200 & GDS-300 Series User Manual 134 File Manager Goes to the file manager to set the save-file path or to create new folders. See page
OPERATION 135 Recall Operations The Recall function configures the save file destination, file format and in the case of image files, the on-screen
GDS-200 & GDS-300 Series User Manual 136 Waveform Recall Items From Selects the source waveform. To Selects the destination to recall to.
OPERATION 137 File Format 5. Chooses the file format that is to be recalled. This option will determine whether *.LSF or *.CSV files are shown in t
GDS-200 & GDS-300 Series User Manual 138 Recalling Setups Background Setup files can be recalled from the internal Set (setup) memories or fr
OPERATION 139 File Manager The file manager is used to navigate the internal file system, create new folders or to delete files. Accessing the File
GDS-200 & GDS-300 Series User Manual 14 Accessories Standard Accessories Part number Description Certificate of calibration 82DS-2200E01
GDS-200 & GDS-300 Series User Manual 140 Navigating the File System Background The following diagram will show the file manager functions.
OPERATION 141 Navigation 1. Scroll up and down to view files that are out of view. 2. Select a folder to enter a directory. 3. Press the Back but
GDS-200 & GDS-300 Series User Manual 142 Exit Press Exit to exit from the file system.
OPERATION 143 DMM Mode The DMM function can be used simultaneously with the DSO function. The DMM function is a basic digital multimeter. By defaul
GDS-200 & GDS-300 Series User Manual 144 Second Method: Vertical Orientation: Press and hold the DMM area for a short while to directly bring
OPERATION 145 Multimeter Display Overview The DMM display will show the measurement mode, max, min and current measurement, and a trend plot (verti
GDS-200 & GDS-300 Series User Manual 146 2. The selected mode will be previewed in the lower corner. VAC and VDC keysMultimeter settingsPrev
OPERATION 147 Steps 1. From the multimeter settings choose ACA, ACmA, DCA or DCmA. Ensure that no more than the 600mA is input into the mA port.
GDS-200 & GDS-300 Series User Manual 148 Connection 10A MAXFUSEDVWCOMA mAGDS-220Digital Storage Oscilloscope200 MHz 1GSa/sCOM VWCAT II 600
OPERATION 149 Diode Measurement Back Ground The diode test checks the forward bias characteristics of a diode by running a constant forward bias cu
GETTING STARTED 15 Appearance GDS-200/300 Front Panel VWCOMA mAGDS-220Digital Storage Oscilloscope200 MHz 1GSa/sFUSED10A MAXFUSEDCAT II600mA MAX
GDS-200 & GDS-300 Series User Manual 150 Note Ensure the diode polarity is correct before performing the diode measurement. Example Diode me
OPERATION 151 2. The selected mode will be previewed in the lower corner. Continuity keyMultimeter settingsPreview screen 3. Press anywhere outs
GDS-200 & GDS-300 Series User Manual 152 Temperature Measurement Back Ground The GDS-300 can measure temperature using a thermocouple. To mea
OPERATION 153 3. The selected mode will be previewed in the lower corner. Temperature key/toggle temperature unitMultimeter settingsPreview screen
GDS-200 & GDS-300 Series User Manual 154 EE APPs The GDS-200/300 has a number of handy engineering apps built-in. The following apps are avai
OPERATION 155 Attenuation Calculator Background The attenuation calculator calculates the resistor values of 5 common attenuator circuits. The onl
GDS-200 & GDS-300 Series User Manual 156 5. Press Calculate. This will calculate the exact resistance values required for the desired attenu
OPERATION 157 Note Please see the DMM chapter for details on the different measurement modes and connections. See from page 143. Select Measuremen
GDS-200 & GDS-300 Series User Manual 158 Calculator Background The Calculator function is a convenient scientific calculator especially suite
OPERATION 159 QR Code Link App Background The QR Code Link app displays QR codes for commonly used websites relating to the GDS-200/300. The GW Ins
GDS-200 & GDS-300 Series User Manual 16 DMM Ports 10A MAXFUSEDVWCOMA mACAT II 600VCAT III 300VFUSED600mA MAX mA Accepts up to 600mA. Fuse pr
GDS-200 & GDS-300 Series User Manual 160 2. Press the Color Code button for standard pin-type resistors. 3. Select the 4 band or 5 band ta
OPERATION 161 SMD 3-digit/4-digit code EIA-96 2%, 5%, 10% 4. Press the SMD code button and enter the SMD code in the pop-up window. 5. Press
GDS-200 & GDS-300 Series User Manual 162 Example Source templateTemplateTemplate offset from source groundEdit window Steps 1. From the EE Ap
OPERATION 163 left and right to position the green edit window and press the Edit button. Each point within the edit window can now be edited to
GDS-200 & GDS-300 Series User Manual 164 REMOTE CONTROL CONFIG This chapter describes basic configuration for remote control. For a command li
REMOTE CONTROL CONFIG 165 Interface Configuration The GDS-200/300 uses the USB device port for remote control. When using the remote control functio
GDS-200 & GDS-300 Series User Manual 166 Remote Control Function Check Functionality check Invoke a terminal application such as Realterm. I
REMOTE CONTROL CONFIG 167 Using Realterm to Establish a Remote Connection Background Realterm is a terminal program that can be used to communica
GDS-200 & GDS-300 Series User Manual 168 The baud rate, stop bit and parity settings can be viewed by right-clicking the connected device and
REMOTE CONTROL CONFIG 169 5. After Realterm has started, click on the Port tab. Enter the Baud, Parity, Data bits, Stop bits and Port number confi
GETTING STARTED 17 Rear Panel Horizontal standVertical standVent Vent Do not block the vent opening. Vertical Stand Use the vertical stand to posit
GDS-200 & GDS-300 Series User Manual 170 6. Click on the Send tab. In the EOL configuration, check on the +CR and +LF check boxes. Enter the
FAQ 171 FAQ • I connected the signal but it does not appear on the display. • I want to remove the measurement or FFT results from the display. •
GDS-200 & GDS-300 Series User Manual 172 The waveform does not update (frozen). Press the Run/Stop key to unfreeze the waveform. See page 80
APPENDIX 173 APPENDIX The Appendix contains basic maintenance, miscellaneous information and the device specifications. Firmware Update ...
GDS-200 & GDS-300 Series User Manual 174 DMM Specifications ...
APPENDIX 175 Firmware Update Background The GDS-200/GDS-300 firmware can be upgraded by copying the new firmware to the internal memory and install
GDS-200 & GDS-300 Series User Manual 176 Fuse Replacement Background The GDS-200/300 has fuse protection for both current ports, a 10A and a
APPENDIX 177 Resistor Standards Resistor Standards Standard Tolerance Standard Tolerance E6 20% E48 2% E12 10% E96 1% E24 5% E192 0.5% Standard Res
GDS-200 & GDS-300 Series User Manual 178 3 or 4 Digit SMD Resistor Standards For 3 or 4 digit SMD codes, the first 2 or 3 digits represent th
APPENDIX 179 Multiplier Letter Code Multiplier Letter Code Multiplier F 105 B 101 E 104 A 100 D 103 X or S 10-1 C 102 Y or R 10-2 Example: 01A = 10
GDS-200 & GDS-300 Series User Manual 18 Interface Panel Power input for optionsCalibration outputPower switchInternal use ports (restricted ac
GDS-200 & GDS-300 Series User Manual 180 GDS-200/300 Default Settings The following default settings are the factory configuration settings fo
APPENDIX 181 GDS-200/300 Specifications The specifications apply when the GDS-200/GDS-300 is powered on for at least 30 minutes. Scope Specificatio
GDS-200 & GDS-300 Series User Manual 182 Sensitivity DC - 25MHz: approx. 0.5div or 5mV 25MHz - 70/100/200MHz: approx. 1.5div or 15mV Horizonta
APPENDIX 183 Display Type 7 Inch Display Resolution 480 x 800 Display Orientation Landscape and Portrait Backlight Control Manual adjustable &
GDS-200 & GDS-300 Series User Manual 184 AC Voltage Range 50mV, 500mV, 5V, 50V, 700V 5 ranges Accuracy 50mV, 500mV, 5V, 50V, 700V ±(1.5% + 1
APPENDIX 185 GDS-200/300 Dimensions 10A MAXFUSEDVWCOMA mA59.7136.0240.2GDS-220Digital Storage Oscilloscope200 MHz 1GSa/sMW16.5pF300Vrms MAX.11
GDS-200 & GDS-300 Series User Manual 186 Declaration of Conformity We GOOD WILL INSTRUMENT CO., LTD. No. 7-1, Jhongsing Rd, Tucheng Dist., Ne
INDEX 187 INDEX Accessories ... 14 Advanced DMM ... 156 Amplitude measurements ... 86 Area mea
GDS-200 & GDS-300 Series User Manual 188 Channel selection ... 51 Cursor operation ... 83 Display
GETTING STARTED 19 External Power Port. This port supplies power for optional accessories such as the differential probe option (GW part number: GD
This manual contains proprietary information, which is protected by copyright. All rights are reserved. No part of this manual may be photocopied,
GDS-200 & GDS-300 Series User Manual 20 BNC Panel CH2 inputTapped holeCH1 inputMW16.5pF300Vrms MAX.11 2CAT 300V CH1, CH2 Inputs Channel 1 an
GETTING STARTED 21 Display Overview Horizontal Battery indicator, trigger level, and horizontal positionInput frequency, trigger state.Date, time a
GDS-200 & GDS-300 Series User Manual 22 Vertical In order: Input frequency, trigger state, time, horizontal position, acquisition mode, trigge
GETTING STARTED 23 Horizontal Window Indicator Shows the current horizontal scale and position in relation to the record length. Drop Down Menu Are
GDS-200 & GDS-300 Series User Manual 24 Run Mode (Backward, Run/Stop, Forward, Single) The Run/Stop key is used to start/stop acquiring samp
GETTING STARTED 25 How to Use the Touch Interface Background Unlike other oscilloscopes or digital multimeters, the GDS-200 and GDS-300 Series have
GDS-200 & GDS-300 Series User Manual 26 Swipe Horizontally swipe on the screen. Scroll Drag your finger vertically on the screen to scro
GETTING STARTED 27 Drag and drop DropDrag Drag an item to an option and let go*. *The drag and drop gesture is usually used in the Horizontal, Ver
GDS-200 & GDS-300 Series User Manual 28 Help Menu Background If at any time you are not sure how to use a particular function, the help icon
GETTING STARTED 29 Tap on the Help option in the Utility menu. The help screen overlay is shown on the display. There are three help screens. Use
SAFETY INSTRUCTIONS 3 Table of Contents SAFETY INSTRUCTIONS ... 5 GETTING STARTED ...
GDS-200 & GDS-300 Series User Manual 30 Initial Setup This chapter should be used to set up the instrument when used for the first time or if
GETTING STARTED 31 6. The battery indicator is shown on the top right-hand corner of the main display. USB Driver Installation Background The
GDS-200 & GDS-300 Series User Manual 32 7. The GDS-200/300 will now be recognized as a virtual com port (VCP). Tilting the Stand Horizontal
GETTING STARTED 33 Setting the System Date & Time Background The GDS-200/300 will keep the date and time settings saved so long as the battery
GDS-200 & GDS-300 Series User Manual 34 Setting the Language Background The GDS-200/300 menu system can be set to display a number of langua
GETTING STARTED 35 2. Set the probe attenuation to x10 on the probe. 3. Connect the probe to the 2V calibration signal. The calibration signal out
GDS-200 & GDS-300 Series User Manual 36 interpolation type. 13. Turn the adjust point on the probe until the probe is properly compensated.
OPERATION 37 OPERATION Gesture Control ... 41 Select a Channel, Math or Reference Wavef
GDS-200 & GDS-300 Series User Manual 38 Trigger Parameters and Settings ...
OPERATION 39 Acquire Menu ... 107 Accessing the Acquisition Menu ...
GDS-200 & GDS-300 Series User Manual 4 APPENDIX ... 173 Firmware Update ...
GDS-200 & GDS-300 Series User Manual 40 Current Measurement ...
OPERATION 41 Gesture Control This chapter will go over all of the basic touch gestures that are available using the default settings. The touch gest
GDS-200 & GDS-300 Series User Manual 42 Set the Vertical Position of the Selected Waveform Background The touch screen can be used to conveni
OPERATION 43 Set the Vertical Scale of the Selected Waveform Background The touch screen can be used to conveniently set the vertical scale of the
GDS-200 & GDS-300 Series User Manual 44 Set the Horizontal Position Background The touch screen can be used to conveniently set the horizonta
OPERATION 45 Set the Horizontal Scale Background The touch screen can be used to conveniently set the horizontal scale of the displayed waveforms.
GDS-200 & GDS-300 Series User Manual 46 Select the Trigger Source and the Trigger Level Background The touch screen can be used to convenient
OPERATION 47 Common Menu UI Elements This section will give a brief overview of some of the common UI elements that are used to select parameters an
GDS-200 & GDS-300 Series User Manual 48 Example SliderPreset value button Value Picker Background Value pickers are used to enter discrete va
OPERATION 49 Example: keypad UnitskeypadDeleteEnter When using a keypad: Select the unit Enter the value using the keypad Press Enter to co
SAFETY INSTRUCTIONS 5 SAFETY INSTRUCTIONS This chapter contains important safety instructions that you must follow during operation and storage. R
GDS-200 & GDS-300 Series User Manual 50 Vertical Menu Operation The vertical menu controls the vertical scale and position, channel activation
OPERATION 51 Entering the Vertical Menu Steps 1. Press the vertical icon to bring up the vertical menu. 2. The Vertical Menu appears. Vertical M
GDS-200 & GDS-300 Series User Manual 52 Example Activated channelsSelected/active channel Disabled channels Vertical Scale and Position Backg
OPERATION 53 Scale pickerVertical Position keyPosition sliderDrag and drop the Vertical position key to manually set vertical position Vertical Opt
GDS-200 & GDS-300 Series User Manual 54 Coupling DC, AC, GND Invert Press the Invert toggle switch to invert the waveform for the selected c
OPERATION 55 Note If the vertical scale is changed when the Expand function is set to ground, the signal will expand from the ground level. The grou
GDS-200 & GDS-300 Series User Manual 56 Vertical Options – Reference Waveforms Background The Options menu in the Vertical menu allows you to
OPERATION 57 Setup 1. Select the math channel by selecting the M icon. Page 51 2. Press the Options icon to bring up the math settings. Math w
GDS-200 & GDS-300 Series User Manual 58 Example Math waveform iconMath waveform Math details The Math icon can be expanded to show the sourc
OPERATION 59 Horizontal Position Background The horizontal scale and position can be configured in the Horizontal menu in addition to the using the
GDS-200 & GDS-300 Series User Manual 6 Safety Guidelines General Guideline CAUTION Do not place any heavy object on the instrument. Av
GDS-200 & GDS-300 Series User Manual 60 7. To manually set the horizontal position, drag from the Horizontal Position key and drop to the key
OPERATION 61 By Trigger Position Scales the waveform from the trigger position. Steps 1. Access the Horizontal or Zoom menu. Page 59 or 63 2. P
GDS-200 & GDS-300 Series User Manual 62 Example: By Trigger Position Expand By TriggerTrigger Position
OPERATION 63 Zoom Background The Zoom menu can be accessed from the Horizontal menu. Setup 1. Press the Horizontal key to access the Horizontal m
GDS-200 & GDS-300 Series User Manual 64 Menu Control 1. Use the scale picker to set the horizontal scale for the zoom window. 2. Use the pos
OPERATION 65 2. The horizontal position of the zoom window can be moved by swiping the screen left or right. Zoom horizontal positionZoom window N
GDS-200 & GDS-300 Series User Manual 66 Trigger Menu Operation The Trigger menu controls the trigger position, trigger type and trigger modes.
OPERATION 67 Entering the Trigger Menu Steps 1. Press the Trigger icon to bring up the Trigger menu. 2. The Trigger menu appears. Example Trigg
GDS-200 & GDS-300 Series User Manual 68 Trigger Setting Trigger Types The trigger configures the conditions for when the GDS-200/300 captures
OPERATION 69 Trigger Parameters and Settings The following trigger parameters apply to all the trigger types unless stated otherwise. Mode Auto The
SAFETY INSTRUCTIONS 7 AC Adapter Power Supply (AC Mains) WARNING AC Input voltage range: 100VAC-240VAC Frequency: 47Hz-63Hz Output: 12 DCV
GDS-200 & GDS-300 Series User Manual 70 50% Sets the trigger level to 50% of the waveform amplitude. Source CH1, CH2 Channel 1 or channel 2
OPERATION 71 When Pulse width (10ns - 10s) Standard NTSC National Television System Committee PAL Phase Alternate by Line SECAM SEquential Couleu
GDS-200 & GDS-300 Series User Manual 72 Forced Trigger Background If the input waveforms are not triggered, they can be force triggered usin
OPERATION 73 Trigger Level Background The trigger level can be set using touch gestures or from the trigger level. Touch Control 1. Select the tr
GDS-200 & GDS-300 Series User Manual 74 Edge Trigger Background The Edge trigger allows you to acquire a waveform on a rising or falling edge
OPERATION 75 Example
GDS-200 & GDS-300 Series User Manual 76 Pulse Trigger Background The Pulse trigger allows you to acquire a waveform on a rising or falling pu
OPERATION 77 Example Video Trigger Background The Video trigger allows you to acquire PAL, NTSC and SECAM video signals. The GDS-200/300 can trigg
GDS-200 & GDS-300 Series User Manual 78 Source CH1, CH2 Standard NTSC, PAL, SECAM Polarity Positive, negative Trigger On Field 1 (lin
OPERATION 79 AC Priority AC Priority mode will scale the waveform to the screen by removing any DC component. Steps 1. From the main display pres
GDS-200 & GDS-300 Series User Manual 8 Storage environment Location: Indoor Temperature: -25°C to 70°C Relative Humidity: <90% Dis
GDS-200 & GDS-300 Series User Manual 80 Run Mode The scope has three main running modes, Run, Stop and Single. Run mode continuously captures
OPERATION 81 Run/Stop Mode Background By default the scope is in Run mode as soon as it is turned on. The Run/Stop key is shown in green when the
GDS-200 & GDS-300 Series User Manual 82 Example Sample sliderSample numberTrigger status Single Mode Single Mode Simply press the Single key
OPERATION 83 Cursor Measurements Horizontal or vertical cursors are used to show the position and values of waveform measurements and math operation
GDS-200 & GDS-300 Series User Manual 84 4. Cursors will now appear on the display for the currently selected channel, 2 lined cursors for th
OPERATION 85 6. Pressing the cursor window will toggle the horizontal reading between frequency and time. Cursor Reading Moving the Cursors 7. T
GDS-200 & GDS-300 Series User Manual 86 Automatic Measurements Up to 4 automatic measurements can be calculated from the trigger configuration
OPERATION 87 Cycle RMS The root mean square value is calculated for all data samples within the first cycle found in the gated region. Area Measur
GDS-200 & GDS-300 Series User Manual 88 –Width Negative pulse width. Duty Cycle Ratio of signal pulse compared with whole cycle. =100x (Puls
OPERATION 89 LFR Time between: Source 1 first falling edge and Source 2 last rising edge. LFF Time between: Source 1 first falling edge and Source
SAFETY INSTRUCTIONS 9 Power cord for the United Kingdom When using the instrument in the United Kingdom, make sure the power cord meets the followin
GDS-200 & GDS-300 Series User Manual 90 Steps 1. Press the M (Measure) icon in the menu bar at the bottom of the display. 2. The minimized v
OPERATION 91 Steps 1. Make sure the automatic measurements are visible at the bottom of the display. 2. From the Menu Bar: Long-press anywhere wi
GDS-200 & GDS-300 Series User Manual 92 5. Select an automatic measurement from the selected category. Tip: Tapping the bottom of the select
OPERATION 93 2. Press the Options icon to bring up the options menu. 3. From the Options menu select the measurement gating: Gating Off (Full
GDS-200 & GDS-300 Series User Manual 94 Drop-Down Menu The Drop-down menu is split into the Menu tray and the App tray. The Menu tray contains
OPERATION 95 Steps Press the Default icon from the Menu tray. The default settings will be restored immediately. Panel Lock Background The panel
GDS-200 & GDS-300 Series User Manual 96 Utility Menu The Utility Menu configures a number of general system settings such as the System langua
OPERATION 97 View System Information Background The system information displays the model name, serial number, firmware version, FPGA revision and
GDS-200 & GDS-300 Series User Manual 98 Signal Path Compensation Background Signal Path Compensation (SPC) is used to compensate the internal
OPERATION 99 Setting the System Date & Time Background The GDS-200/300 will keep the date and time settings saved so long as the battery is no