Gw-instek GPI-700A Series Programming Manual Instrukcja Użytkownika Strona 25

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 31
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 24
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
47
:IRESistance:TTIMe
Function:
Set or query the IR test time.
Syntax:
:IRESistance:TTIMe <NR2>
:IRESistance:TTIMe?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:IRESistance:TTIMe 5.0 set the IR test time to 5 sec.
:IRESistance:TTIMe? return 005.0 if the IR test time setting is 5 sec.
:GBONd:CURRent
Function:
Set or query the GB test current.
Syntax:
:GBONd:CURRent <NR1>
:GBONd:CURRent?
<NR1> Please refer to the specification
Return:
<NR1>
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
48
Examples:
:GBONd:CURRent 25.00 set the GB test current to 25A.
:GBONd:CURRent? return 25.00 if the GB test current setting is 25A.
:GBONd:RMINimum
Function:
Set or query the GB test minimum resistance.
Syntax:
:GBONd:RMINimum <NR1>
:GBONd:RMINimum?
<NR1> Please refer to the specification
Return:
<NR1>
Examples:
:GBONd:RMINimum 100.0 set the GB test minimum resistance to 100mΩ.
:GBONd:RMINimum? return 100.0 if the GB test minimum resistance setting is
100mΩ.
:GBONd:RMAXimum
Function:
Set or query the GB test maximum resistance.