GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
37
:DCContinue:VOLTage?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCContinue:VOLTage 1.000 set the CDC test voltage to 1.000kV.
:DCContinue:VOLTage? Return 1.000 if the CDC test voltage setting is 1.000kV.
:DCContinue:CMINimum
Function:
Set or query the CDC test minimum current.
Syntax:
:DCContinue:CMINimum <NR2>
:DCContinue:CMINimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCContinue:CMINimum 1.00 set the CDC test minimum current to 1.00mA.
:DCContinue:CMINimum? return 01.00 if the CDC test minimum current setting
is 1.00mA.
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
38
:DCContinue:CMAXimum
Function:
Set or query the CDC test maximum current.
Syntax:
:DCContinue:CMAXimum <NR2>
:DCContinue:CMAXimum?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:DCContinue:CMAXimum 1.00 set the CDC test maximum current to 1.00mA.
:DCContinue:CMAXimum? return 01.00 if the CDC test maximum current
setting is 1.00mA.
:DCContinue:RTIMe
Function:
Set or query the CDC ramp time.
Syntax:
:DCContinue:RTIMe <NR2>
:DCContinue: RTIMe?
<NR2> Please refer to the specification
Return:
<NR2>