GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
27
:FUNCtion:TEST:MODe? return 3 if the test function setting is IR.
:FUNCtion:TEST:STATe
Function:
Set or query the test state (0. OFF 1. ON 2. CONTINUE). During testing, when
proceed HOLD function, the instrument will stop testing until a command of
“FUNC:TEST:STAT 2” has been received.
Syntax:
:FUNCtion:TEST:STATe <NR1>
:FUNCtion:TEST:STATe?
<NR1> is in the range from 0 through 1.
Return:
<NR1>
Examples:
:FUNCtion:TEST:STATe 1 start test.
:FUNCtion:TEST:STATe ? return 1 if the test state is 1.
:MEASure
Function:
Return measurement value
Syntax:
:MEASure<X>?
<X> is in the range from 1 through 16 for STEP.
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
28
: MEASure? Return measurement value of COM mode.
Return:
Status, Mode, V/I, Meas, Time
(2 bytes) (1 byte) (7 bytes) (8 bytes) (7 bytes)
Status: 00:Emtpy
10:Pass
20:Fail
21:Interlock
22:Low Voltage
23:Hi voltage
24:ARC
25:Measurement over limit maximum
26:Measurement below limit minimum
30:Stop
32:Hight Voltage
33:Heatsink is too hot
Mode: 0: Empty
1:ACW
2:DCW
3:IR
4:Cnt (GB for 745A)
5:CAC
6:CDC
7:HOLD