GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
41
:ACWStanding:RTIMe
Function:
Set or query the ACW ramp time.
Syntax:
:ACWStamding:RTIMe <NR2>
:ACWStamding:RTIMe?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
:ACWStanding:RTIMe 5.0 set the ACW ramp time to 5 sec.
:ACWStanding:RTIMe? return 005.0 if the ACW ramp time setting is 5 sec.
:ACWStanding:TTIMe
Function:
Set or query the ACW test time.
Syntax:
:ACWStamding:TTIMe <NR2>
:ACWStamding:TTIMe?
<NR2> Please refer to the specification
Return:
<NR2>
GPI/GPT SERIES PROGRAMMABLE EST
PROGRAMMER MANUAL
42
Examples:
:ACWStanding:TTIMe 5.0 set the ACW test time to 5 sec.
:ACWStanding:TTIMe? return 005.0 if the ACW test time setting is 5
sec.
:DCWStanding:VOLTage
Function:
Set or query the DCW test voltage.
Syntax:
:DCWStanding:VOLTage <NR2>
:DCWStanding:VOLTage?
<NR2> Please refer to the specification
Return:
<NR2>
Examples:
: DCWStanding:VOLTage 1.000 set the DCW test voltage to 1.000kV.
: DCWStanding:VOLTage? return 1.000 if the DCW test voltage setting is
1.000kV.
:DCWStanding:CMINimum
Function:
Set or query the DCW test minimum current.
Syntax: