Gw-instek GDS-3000 Series Programming Manual Instrukcja Użytkownika Strona 61

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 103
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 60
COMMAND DETAILS
61
:MEASure:FREQuency
Query
Description
Returns the frequency value.
Syntax
:MEASure:FREQuency{?}
Return parameter
Returns the frequency as <NR3>.
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH1
:MEASure:FREQuency?
1.0E+3
Selects Channel 1, and then measures the
frequency.
:MEASure:NWIDth
Query
Description
Returns the first negative pulse width timing.
Syntax
:MEASure:NWIDth{?}
Return parameter
Returns the negative pulse width as <NR3>
Note
Before using this command, select the
measurement channel. See the example below.
Example
:MEASure:SOURce1 CH 1
:MEASure:NWIDth?
Selects Channel 1, and then measures the negative
pulse width.
:MEASure:PDUTy
Query
Description
Returns the positive duty cycle ratio as percentage.
Syntax
:MEASure:PDUTy{?}
Return parameter
<NR3>
Przeglądanie stron 60
1 2 ... 56 57 58 59 60 61 62 63 64 65 66 ... 102 103

Komentarze do niniejszej Instrukcji

Brak uwag