Gw-instek GDS-3000 Series Programming Manual Instrukcja Użytkownika Strona 102

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 103
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 101
GDS-3000 Programming Manual
102
Parameter
<NR2>
Desired template position (-12.0 ~ +12.0
divisions)
Return parameter
Returns the position in the following format:
“<NR2>Div”
Example
:TEMPlate:POSition:MINimum 3.00
Sets the minimum template position to 3.00
divisions.
:TEMPlate:SAVe:MAXimum
Set
Description
Saves the maximum template.
Syntax
:TEMPlate:SAVe:MAXimum
:TEMPlate:SAVe:MINimum
Set
Description
Saves the maximum template.
Syntax
:TEMPlate:SAVe:MINimum
:TEMPlate:TOLerance
Set
Query
Description
Sets or queries the tolerance as a percentage.
Syntax
:TEMPlate:TOLerance{NR2|?}
Parameter/
Return Parameter
<NR2>
The auto tolerance range (0.4% ~ 40%)
Example
:TEMPlate:TOLerance 10
Sets the tolerance to 10%.
:TEMPlate:SAVe:AUTo
Set
Description
Saves the AUTO template (maximum and
minimum templates).
Syntax
:TEMPlate:SAVe:AUTo