GDS-2000A Programming Manual
252
Return parameter
Returns the offset of the function generator
function in <NR3> format.
Example
:FUNCtion1:OFFSet 1.00
Sets the function generator offset to 1.00 volts.
:FUNCtion<x>:DUTYcycle
Description
Sets or queries the duty cycle for the square mode
waveform.
Syntax
:FUNCtion<x>:DUTYcycle {<NRf>|?}
Related
commands
:FUNCtion<x>:MODe
Parameter
<x>
Module 1 or 2.
<NRf>
(5 ~ 95) The duty cycle as a
percentage (5% ~ 95%)
Return parameter
Returns the offset of the function generator
function in <NR3> format.
Example
:FUNCtion1:DUTYcycle 10
Sets the function generator duty cycle to 10%.
Set
Query