The number of characters in the string
depends on the data size setting. The
string must be enclosed in double
quotes, "string".
:TRIGger:BUS:B1:UART:CONDition TXDATA
:TRIGger:BUS:B1:UART:TX:DATa:SIZe 1
:TRIGger:BUS:B1:UART:TX:DATa:VALue "1x1x0101"
Sets the value to 1x1x0101
:TRIGger:BUS:B1:UART:TX:DATa:VALue?
1X1X0101
:TRIGger:BUS:B1:SPI:CONDition
Sets or queries the SPI triggering condition.
:TRIGger:BUS:B1:SPI:CONDition {SS | MISO | MOSI |
MISOMOSI | ? }
Set to trigger on the Slave Selection
condition.
Set to trigger on the Master-In
Slave-Out condition.
Set to trigger on the Master-Out
Slave-In condition.
Set to trigger on the Master-In
Slave-Out and Master-Out Slave-In
conditions.
Returns the triggering condition.
:TRIGger:BUS:B1:SPI:CONDition MISO