COMMAND DETAILS
59
:GONogo:NGCount?
Query
Description
Returns the test result count ratio (failed count,
total count).
Syntax
< Long >
< Short >
:GON:NGC?
:GON:NGC?
Return parameter
<NR1>, <NR1>
<failed count>,<total count>
Example
:GON:NGC?
>2,128
2 fails from 128 Go-
NoGo tests.
:GONogo:NGDefine
Set
Query
Description
Sets or queries the Go-NoGo boundary template
conditions.
Note
Before any Go-NoGo command can be used,
please use the :GONogo:FUNCtion 1 command to
initialize the oscilloscope.
Syntax
< Long >
< Short >
:GONogo:NGDefine {0|1}
:GONogo:NGDefine?
:GON:NGD {0|1}
:GON:NGD
Parameter/
Return parameter
0
No-Go when the waveform doesn’t
exceed the boundary template.
1
No-Go when the waveform exceeds the
boundary template.
Example
:GON:NGD 1
NoGo conditions set
to when outside
template.
:GONogo:SOURce
Set
Query
Description
Sets the Go-NoGo channel source.