COMMAND DETAILS
29
:CHANnel<X>:PROBe:RATio
Set
Query
Description
Sets or returns the probe attenuation factor.
Same as: Channel key variable knob
Syntax
< Long >
< Short >
:channel<X>:probe:ratio<NRf>
<NRf>
:channel<X>:probe:ratio?
:chan<X>:prob:rat
<NRf>
:chan<X>:prob:rat?
Parameter
<X>
Channel
<NRf>
Probe
attenuation
factor
1/2
CH1/2
0.1/0.2/0.5
0.1x/0.2x/0.5x
1/2/5
1x/2x/5x
10/20/50
10x/20x/50x
100/200/500
100x/200x/500x
1000/2000
1000x/2000x
Example
:channel1:probe:ratio 1
Sets the Channel 1
probe attenuation
factor to 1x
:CHANnel<X>:PROBe:TYPE
Set
Query
Description
Sets or returns the probe type (voltage/current).
Same as: Channel key F4
Syntax
< Long >
< Short >
:channel<X>:probe:type
<boolean>
:channel<X>:probe:type?
:chan<X>:prob:type
<boolean>
:chan<X>:prob:type?