[SOURce:]CURRent[:LEVel]:TRIGgered
[:AMPLitude]
Set
Query
Description
Sets or queries the current level in amps when a
software trigger has been generated.
Syntax
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]
{<NRf>|MIN|MAX}
Query Syntax
[SOURce:]CURRent[:LEVel]:TRIGgered[:AMPLitude]?
[MIN|MAX]
Parameter/Return
<NRf>
0%~105% of the rated current output in amps.
MIN
Minimum current level.
MAX
Maximum current level.
Example
SOUR:CURR:LEV:TRIG:AMPL? MAX
37.800
Returns the maximum possible current level in amps.
[SOURce:]CURRent:PROTection[:LEVel]
Set
Query
Description
Sets or queries the OCP (over-current protection)
level in amps.
Syntax
[SOURce:]CURRent:PROTection[:LEVel]
{<NRf>|MIN|MAX}
Query Syntax
[SOURce:]CURRent:PROTection[:LEVel]? [MIN|MAX]
Parameter/Return
<NRf>
OCP range in Amps.
MIN
Minimum current level.
MAX
Maximum current level.
Example
SOUR:CURR:PROT:LEV? MIN
+3.600
Returns the minimum possible current level in amps.