COMMAND DETAILS
21
±5.0
5.0V ~ +5.0V
(100mV/div~500mV/div)
±50.0
50.0V ~ +50.0V
(1V/div ~ 5V/div)
Example
:channel1:scale 1.00e2
:channel1:offset 2.00e2
Sets the Channel 1
scale to 10mV/div
Sets the Channel 1
offset to 20mV
:CHANnel<X>:PROBe
Set
Query
Description
Sets or returns the probe attenuation factor.
Same as: Channel key F4
Syntax
< Long >
< Short >
:channel<X>:probe <NR3>
:channel<X>:probe?
:chan<X>:prob <NR1>
:chan<X>:prob?
Parameter
<X>
Channel
<NR1>
Probe attenuation factor
1/2
CH1/2
0
1x
1
10x
2
100x
Example
:channel1:probe 1
Sets the Channel 1
probe attenuation
factor to 10x
:CHANnel<X>:SCALe
Set
Query
Description
Sets or returns the vertical scale. The scale depends
on the probe attenuation factor.
Same as: Volts/Div knob
Syntax
< Long >
< Short >
:channel<X>:scale <NR3>
:channel<X>:scale?
:chan<X>:scal <NR3>
:chan<X>:scal?