DATA LOGGING REMOTE CONTROL
13
1. Press the Record SetupStart icon. This icon will
only be available after a measurement has been
selected with the Setup Multimeter icon (page 11).
2. Choose the Logging tab.
3. Set the logging start time in Begin Logging Data
panel.
Immediately: Choose this option to begin
logging measurements as soon as the “Play”
icon is pressed. (see page 17 for details)
At Time: This option allows you to choose
the exact time that the logging will start.
The start date, hour, minute and second can
be chosen.
On Receiving External Trigger: This option
will only start logging after the TRIG key is
pressed on the DMM front panel.