Gw-instek GDM-8246 Programming Manual Instrukcja Użytkownika Strona 18

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
  • Strona
    / 28
  • Spis treści
  • BOOKMARKI
  • Oceniono. / 5. Na podstawie oceny klientów
Przeglądanie stron 17
GDM-8246 MULTIMETER
PROGRAMMER MANUAL
34
:READ?(query only)
Function:
Returns the value displayed on the primary and secondary display with
14 characters totally.
Syntax:
:READ?
Examples:
:READ? Might return “ NONE ,+0.0000” when the current function is in
the DC voltage.
:VALUE?(query only)
Function:
Returns the value displayed on the primary and secondary display with 7
characters totally.
Syntax:
:VALUE?
Examples:
:VALUE? might return “+0.0000” when the current function is in the DC
voltage.
:SVALue?(query only)
Function:
Returns the value displayed on the primary and secondary display with 6
characters totally.
GDM-8246 MULTIMETER
PROGRAMMER MANUAL
35
Syntax:
:SVALue?
Examples:
:SVALue? might return “ -OL- ” when the current function is in the dBm
mode.
:CALCulation:LIMit:STATe
Function:
Returns or sets the compare mode on or off.
Syntax:
:CALCulation:LIMit:STATe?
:CALCulation:LIMit:STATe <Boolean>
<Boolean> can be o(off) or 1(on)
Returns:
0/1
Examples:
:CALCulation:LIMit:STATe? might return 1 in the compare mode.
:CALCulation:LIMit:STATe 1 sets the compare mode.
:CALCulation:LIMit:LOWer
Function:
Returns or sets the value of the lower limit in the compare mode.
Syntax:
Przeglądanie stron 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 27 28

Komentarze do niniejszej Instrukcji

Brak uwag